MRI ABDOMAN ARTIFACT REDUCTION BY FAT SATURATION

Fat Suppression 
Fat signal causes high signal respiration induced ghosts on T1 weighted images and T2 weighted Turbo Spin Echo images. Basic fat suppression methods are used in different combinations by different equipment so take time to learn how your scanner suppresses fat signal. These notes only give a basic outline.

    Selective Spectral Saturation
    • aka Chemsat, FATSAT, CHESS spectral non spatial selective saturation

    • A fat selective excitation pulse is applied to the whole volume, followed by a spoiler gradient , Only water spins contribute signal to the following imaging sequence.
    • Usually applied before each excitation pulse (TR x slice)
    • Takes 10 - 20 msec per routine
    • Fat saturation effect lasts about 100 msec
    • Increases SAR
    • Sensitive to magnetic field distortions
    • Only modifies appearance of Fat other contrasts remain unaltered
    • Suitable for use after Gadolinium contrast agents
    • More difficult (and slower) to achieve at lower field strengths.
    Binomial Excitation
    • aka Water Excitation, Jump Back excitation, 1331 excitation
    • Relies on the fat and water spins moving out of phase. A series of broadband low flip pulses are timed to decrease the flip angle of Fat, but increase the flip angle of water spins.
    • Fat Suppression by selective excitation of water (WE)
    • Less susceptible to magnetic field inhomogeneity than spectral saturation
    • Lower SAR loading than FATSAT or STIR
    • Fastest fat suppression routine
    STIR
    • Short TI (tau) Inversion Recovery
    • Fat suppression is based on T1 behaviour and selection of TI
    • Reverse T1 contrast plus T2 contrast
    • Uniform fat suppression independent of magnetic field inhomogeneity
    • Can be implemented at any field strength with equal success
    • Works well with many acquisition regimes
    • Don't use STIR post Gd contrast
    • Higher SAR than FATSAT or Water Excitation
    SPIR
    • aka Spectral Inversion Recovery
    • Spectrally selective form of STIR
    • A spectrally selective pulse is applied (non spatial) with a flip selected between 90 and 180 degrees. After a suitable delay time (TD depending on flip angle) the fat spins have reached Mz=0 and the imaging sequence is run.
    • Compatible with Gadolinium contrast
    • Less susceptible to magnetic field inhomogoneities
    • Must be applied at for each excitation routine (TR x slice)

1 comment:

  1. Herpes has been one of the most significant virus in the US now and its spreading really fast, and the government are only producing medical drugs that can suppress it but rather keep on eliminating the African herbal doctors who were able to discover a way to completely cure the virus and other diseases... I am Dr. Brain .Francisco. I am a female detox herbalist, I prepare herbal medicine from natural roots and herbs that Cures any kind of Disease or Virus like Herpes Virus, Cancer, Arthritis, Blood Pressure, PCOS, Heart Problem, autoimmune disease, ETC. Using Pure natural Roots and Herbs..I’ve been helping people get cured for 23years now and I know the right herbs to use. I don’t know if you may be interested in my herbal products or are you having any virus or diseases in your body system that needs to be cured with my herbal medicine?...or you have someone that’s going through such? You can also reach me on my email.. franciscobrain8@gmail.com,