- Contraindications include the presence of a cardiac pacemaker, cardiac defibrillator, and Swan-Ganz catheter. MR study is also recommended to be delayed for 6 to 8 weeks after an implant or coronary stent placement.
- The cardiac MRI for functional evaluation alone does not require Gd injection. If the patient has a mass, ischemic heart disease or congenital heart disease, then he/she will require Gd injection. If necessary for the further evaluation of cardiopulmonary structures, start an iv line (20 or 22 gauge).
- ECG gating is essential for cardiac imaging. First, prepare the skin for lead placement. If necessary, shave the body hairs for maximum lead-skin contact and scrub the skin with an abrasive to create edema to improve electrical conductivity. Position the ECG electrodes (RA, LA, RL and LL) on the chest of the patient (in supine position, the most motionless area is the posterior chest). It is best to use MRI electrodes which have no metal (only carbon snaps) and a large surface area with abundant lubrication to minimize burns. Do not place leads over bony areas and avoid looping ECG leads which may result in superficial burns and can increase electrical interference from gradient activity. Leads should not be too far away from each other. Click ‘Gating control’ from the Rx manager area, turn on the advanced ECG gating. View R waves for at least 16 heart cycles while patient is lying down still. If there is a difficulty in detecting the R waves, or the R wave amplitude is low (<1mV) reposition the ECG leads. The detected vectors will be I, II, or III leads. Choose the one which has a clean trace. Whenever you change the lead, do not forget to update the R-peak amplitude. If some of the R waves are missing trigger marks, reduce ‘Cardiac Trigger Level’ to 50%.
- Both fast breath-held and respiratory triggered sequences are used in cardiac studies. For respiratory gating and triggering, place the respiratory monitoring bellows around the patient’s abdomen or chest, so the operator will know if the patient is cooperating with breathing instructions.
- Valium (5-10mg po) or Xanax (1-2 mg po) if patient is claustrophobic.
- Provide ear plugs or music headphones.
Patient Preparation of the CARDIAC MRI
Subscribe to:
Post Comments (Atom)
Popular Posts
-
MP-RAGE stands for Magnetisation Prepared RApid Gradient Echo. MP-RAGE is a 3DFT contrast prepared gradient echo sequence. The T1 weig...
-
High Resolution Brainstem Imaging CISS (Constructive Interference Steady State) CISS is a strongly T2 weighted GRE sequence. In essen...
-
Normal MRI Anatomy The pectoralis tendon is best seen on axial T1 and T2-weighted images as a curvilinear low-signal band inserting o...
-
Because MR neurography sequences are relatively lengthy and motion sensitive, it is not generally feasible to screen the whole extremity ...
-
Administer an infusion of SPIO particles over 30 minutes,then wait 30-60 minutes Coronal single shot echo train spin echo sequence Ax...
-
SAGITAL ORBIT MRI PLANNING REFERRAL LINES AXIAL ORBIT MRI PLANNING REFERRAL LINES CORONAL ORBIT MRI PLANNING REFERRAL ...
-
--> Routine shoulder protocols are the most variable, but most include SHOULDER MRI SEQUENCES ...
-
3-PLANE LOC AXIAL T1 AXIAL T2 FS CORONAL T1 CORONAL STIR Axial T2 high resolution FIESTA ( Facial Nerves in IAC for any space occu...
-
The force of the MRI magnet caused the IV cart to lift up and it flew through the air, on to the MRI machine. Patient was lying outside of...
-
Axial Plane: Prescribe plane perpendicular to midshaft of the proximal phalanx of the thumb. Scan from 1st carp-metacarpal joint th...
No comments: